The AutoVent 3000 includes additional inspiratory time settings to suit a wide range of patient applications. This ventilator is perfect for critical emergency care situations.
| Control Modules |
| Supply Pressure Range: |
40-60 psig |
| Operating Temperature: |
0° F to 125° F |
| Storage Temperature: |
-40° F to 160° F |
| Frequency: |
8 to 28 BPM
|
| Tidal Volume: |
200 to 1200ml
|
| Flow Rate: |
16 to 48 lpm |
| Inspiratory Time: |
approx .75, 1.5 or 2.0 seconds |
| Expiratory Time |
1.5 to 6.0 seconds |
| I-E Ratio: |
1:1 to 1:4 |
| Weight: |
24 oz. (680 grams)
|
| Dimension: |
Width: 6"
Depth: 1.75"
Height: 3.5"
|
| Case Material: |
Polyester |
| Input Connection |
plated brass
|
| Output Connection: |
plated brass |
| Minute Volume: |
0 to 24 lpm |
|
Gas Consumption
Driving Gas:
|
0.5 lpm maximum |
| Driving Gas: |
0.5 lpm maximum |
| Patient Valve Assembly |
| Flow: |
as required in demande mode: 0-48 lpm at 50 psig |
| Peak Inspiratory Flow: |
48 lpm at an airway pressure drop of less than 2.5 cm H20 |
| Delivery Pressure: |
60 +-5cm H20 |
| (Insp.) Crack Pressure: |
0 to -2 cm H20 (0 to -.8 in. H20) |
| Exhalation Resistance: |
LPM
0-10
11-70 |
cm H20 (max)
1.5
3.8 |
| Gas Consumption Driving Gas: |
0.5 lpm maximum |
| Dead Space: |
8 ml (excluding mask) |
| Supply Pressure: |
40 to 60 psig |
| Operating Temperature: |
0 F to 125 F |
| Storage Temperature: |
-40 F to 160 F |
| Inlet Fitting: |
standard male oxygen DISS |
| Filter: |
25 micron stainless steel mesh |
| Outlet: |
22 mm outside diameter x 15 mm inside diameter; fits standard medical masks, endotracheal tubes, and tracheostomy tubes |
| Weight |
16 oz. (450 grams) |
| Materials: |
body - anodized aluminum; cover - polycarbonate; outlet fitting - polycarbonate; inlet fitting - plated brass |