Airon MACS CPAP System

Airon MACS CPAP System for EMS and Hospital Use


Airon MACS CPAP System for EMS and Hospital Use

Product Code: 20201


Call to Order - 1-800-226-3783 ViewDetails



The MACS CPAP system is a face mask device for EMS and Hospital Use. This CPAP is designed for application by mask or endotracheal tube. Indications include the following: Congestive Heart Failure (CHF), Asthma, Acute Respiratory Failure (ARF), Trauma, COPD, Near Drowning, or patients with severe difficulty breathing.  Same great performance as the CPAP system in the pNeuton ventilator.  Also uses the same patient circuits 58001 & 58051.

Controls: CPAP: from 0 to 20 cm H20
Operating Ranges: Internal P Limit: 42 cm H20
Accuracy of Controls: CPAP +-5%
FIO2 +- 10%
Precision (breath to breath repeatability of controls): CPAP +-2 CM H20
FIO2 +-5%
Specificity (effect of one control on another): CPAP +- 5%
FIO2 +- 5%
Internal Compliance: 0.1 ml/cm H20
Device Resistance to Flow: Inspiratory, 60 l/min: 2 cm H20/I/sec
Expiratory, 50 l/min: 2 cm H20/I/sec
Hypobaric Compatibility: up to 15,000 feet (5,000 meters)
Weight: 3 pounds (1.35 kg)
Size:

Width: 3.5" (9cm)
Depth: 8" (20cm)
Height: 6.75" (17cm)

Storage Temp. Range: -46 to 71°C (-51 to 160°F),
15 to 95% humidity noncondensing
Operating Temp. Range: -26 to 60°C (-15 to 140°F),
15 to 95% humidity noncondensing
Driving Gas Requirement: 55 psi +- 15 psi (380 kPa)
100% O2 Do not use the device with other types of gases.
The Gas supply must be capable of delivering at least 40 liters per minute at 55 psi. If input pressure drops below 40 psi due to insufficient gas flow, the device will begin to malfunction.