Home | About Us | Register |   Login | My Account    
0 Items  |   $0.00  |   Checkout
Home » Limited Territory Products » Airon pNeuton Transport Ventilator Model S

Airon pNeuton Transport Ventilator Model S

Airon pNeuton Transport Ventilator Model S


Airon pNeuton Transport Ventilator Model S

Product Code: 20002


Call to Order - 1-800-226-3783 ViewDetails



The pNeuton model S Ventilator is a high quality, lightweight transport ventilator weighing only 6 pounds. Operates completely pneumatically without the need for batteries or external power. This ventilator is also MRI compatible up to 1.5T. Also includes CPAP function.  This ventilator can care for the most difficult patients wherever they may be located.

Accessories

Controls: Mandatory Breaths: On or Off
Respiratory Rate: from 2 to 50 bpm
Tidal Volume: from 360 to 1,500 ml
Peak Pressure: from 10 to 75 cm H20
% Oxygen: 100% or 65%
Operating Ranges: Inspiratory Time: 0.6 to 2.5 seconds
Expiratory Time: 0.6 to 30.0 seconds
Minute Volume: 0.2 to 30 L/min
Internal P Limit: 80 cm H20
Accuracy of Controls: Respiratory Rate: +- 10% (VT between 500-900)
Tidal Volume +- 10%
Peak Pressure +- 10%
PEEP/CPAP +- 5%
FIO2 mandatory breaths +- 10%
Precision (breath to breath repeatability of controls):

Respiratory Rate: +- 10%
Tidal Volume: +- 25ml
Peak Pressure: +- 5 cm H20
PEEP/CPAP: +- 2 cm H20
FIO2 +- 5%

Specificity (effect of one control on another):

Respiratory Rate: if tidal volume is constant, +- 5%.
Tidal Volume: +-5%
Peak Pressure: +- 5%
PEEP/CPAP: +- 5%
FIO2: +- 5%

Internal Compliance: 0.1 ml/cm H20
Device Resistance to Flow: Inspiratory, 60 l/min: 2 cm H20/l/sec
Expiratory, 50 l/min: 2 cm H20/l/sec
Hypobaric Compatibility: Compatible up to 15,000 feet
(5,000 meters)
Weight: 6 pounds (2.7 kg)
Size:

Height: 7" (17cm)
Width: 10" (25cm)
Depth: 7" (17cm)

Storage Temp. Range: -20 to 60°C (-5 to 140°F),
15 to 95 percent humidity, noncondensing
Operating Temp. Range: -5 to 40°C (23 to 105°F),
15 to 95 percent humidity, noncondensing
Driving Gas Requirement:

55 psi +- 15 psi (380 kPa +- 100 kPa)
100% O2. Do not use the ventilator with other types of gases.
The gas supply must be capable of delivering at least 40 liters per minute at 55 psi. If input pressure drops less than 40 psi due to insufficient gas flow, the ventilator will alarm and begin to malfunction.

MRI Compatible with Scanners up to: Maximum static field strength - 1.5 T
Maximum spatial field gradient - 260 G/cm
RF transmitter power - 250 V