The pNeuton model A ventilator is MRI compatible due to it's completely pneumatic operation. MRI Compatible up to 3.0T. Ventilate without batteries or external power. This ventilator includes CPAP function. Choose from noninvasive or invasive ventilation. Includes patient disconnect and low gas alarms. This ventilator has simple to use controls.
| Controls: |
Mandatory Breaths: On or Off
Respiratory Rate: from 3 to >28 bpm
Tidal Volume: from 360 to 1,500 ml
Peak Pressure: from 15 to 75 cm H20
% Oxygen: 100% or 65% |
| Operating Ranges: |
Inspiratory Time: 0.6 to 2.5 seconds
Expiratory Time: 0.6 to 20.0 seconds
Minute Volume: 0.2 to 30 L/min
Internal P Limit: 80 cm H20 |
| Accuracy of Controls: |
Respiratory Rate: +- 10% (VT between 500-900)
Tidal Volume +- 10%
Peak Pressure +- 10%
PEEP/CPAP +- 5%
FIO2 mandatory breaths +- 10% |
| Precision (breath to breath repeatability of controls): |
Respiratory Rate: +- 10%
Tidal Volume: +- 25ml
Peak Pressure: +- 5 cm H20
PEEP/CPAP: +- 2 cm H20
FIO2 +- 5%
|
| Specificity (effect of one control on another): |
Respiratory Rate: if tidal volume is constant, +- 5%.
Tidal Volume: +-5%
Peak Pressure: +- 5%
PEEP/CPAP: +- 5%
FIO2: +- 5%
|
| Internal Compliance: |
0.1 ml/cm H20 |
| Device Resistance to Flow: |
Inspiratory, 60 l/min: 2 cm H20/l/sec
Expiratory, 50 l/min: 2 cm H20/l/sec |
| Hypobaric Compatibility: |
Compatible up to 15,000 feet
(4,600 meters) |
| Weight: |
6.5 pounds (3 kg) |
| Size: |
Height: 5" (13cm)
Width: 10" (25cm)
Depth: 7" (18cm)
|
| Storage Temp. Range: |
-20 to 60°C (-4 to 140°F),
15 to 95 percent humidity, noncondensing |
| Operating Temp. Range: |
-5 to 40°C (23 to 104°F),
15 to 95 percent humidity, noncondensing |
| Driving Gas Requirement: |
55 psi +- 15 psi (380 kPa +- 100 kPa)
100% O2. Do not use the ventilator with other types of gases.
The gas supply must be capable of delivering at least 40 liters per minute at 55 psi. If input pressure drops less than 30 psi due to insufficient gas flow, the ventilator will alarm and begin to malfunction. |
| MRI Compatible with Scanners up to: |
Maximum static field strength - 3 T
Maximum spatial field gradient - 6.9 G/cm
RF transmitter power - 300 V
Shielded |