INTRAPULMONARY
PERCUSSIVE VENTILATION (IPV)
SUGGESTED
VENTILATOR PROTOCOL
VENTILATOR
PATIENT PROCEDURE:
1. For PB 7200 ventilator: Treatment
can be done in either volume or pressure control
A.
Place
ventilator in the SIMV or CPAP Mode.
1. A/C
mode with volume control ventilation will cause each
percussion to be delivered at ventilator settings.
B. Turn off Pressure Support/Flow-By
C.
Respiratory
rate of > 8 breaths per minute (bpm) (7200
ventilator only)
1. On
the 7200, with respiratory rates < 8 bpm, while administering Intrapulmonary
Percussive Ventilation therapy inline, perceived tidal volumes my register at
> 8 L during one breath cycle. This
will cause a 4205-error code. When this error code appears, the 7200 shuts down.
2.
If three of the 4205 alarms occur,
within a 24-hour period, the 7200 stops operation and
requires an EST to be run.
3. Mean
Airway Pressures will increase slightly with the administration of
Intrapulmonary Percussive Ventilation inline with the ventilator. The Respiratory Care Practitioner needs to be
aware of this effect and monitor the patient closely for any adverse side
effects.
D.
PEEP
levels may be maintained.
2. Fill
medication nebulizer with Normal Saline along with prescribed dosage of
respiratory medication if indicated, for a total volume of 20 cc’s. (i.e.; 0.5 cc Albuterol and 3.0 cc Mucomyst and 16.5
cc’s Normal Saline).
3. Place
ventilator Intrapulmonary Percussive Ventilation circuit inline (see drawing)
A. A starting drive pressure of 30 psi is recommended
for ventilated patients.
B. A starting percussive rate of
approximately 200 bpm is recommended.
4. Remote
Switch knob must be activated manually. Percussion should be continued through
two complete ventilator cycles. Release Control knob and allow ventilator to
deliver severa1 machine breaths. Observe visible chest movement. Monitor breath
sounds and observe pulse oximeter for oxygen saturation improvement.
5. After
several percussion intervals lower the endotracheal tube/tracheostomy tube cuff
pressure approximately 1-2 cmH2O to facilitate secretion
mobilization around the ET tube and to prevent possible ET tube obstruction by
secretion impaction.
6. If
chest percussion is inadequate, raise Drive Pressure (PSI gauge) and scan
Frequency Rate to mobilize secretions. Suction as required.
7. Therapy should continue
intermittently for approximately twenty minutes.
8. Patients
who are performing T-tube trials or CPAP sprinting may be taken off of the
ventilator for the Intrapulmonary Percussive Ventilation treatment utilizing an
elbow adapter and 6 inches of aerosol tubing.
Decreasing cuff pressure still applies to this patient population.
Note:
manually cancel occasional high-pressure alarms as they occur. The goal is to achieve approximately twice
the
9. At
Completion of Therapy
A.
Reset cuff pressure.
B.
Remove Intrapulmonary
Percussive Ventilator circuit assembly and store appropriately.
C. Return
ventilator settings if changed prior to starting Intrapulmonary Percussive
Ventilation treatment.
INTRAPULMONARY
PERCUSSIVE VENTILATION (IPV)
SUGGESTED
VENTILATOR PROTOCOL
VENTILATOR
PATIENT PROCEDURE:
1. For PB 840 ventilators, settings
need to be set as follows: Treatment can be done in either volume or
pressure control
A.
In
the SIMV or CPAP Mode.
B.
Turn
off Pressure Support
C.
PEEP
levels may be maintained
1. Mean
Airway Pressures will increase slightly with the administration of
Intrapulmonary Percussive Ventilation inline with the ventilator. The Respiratory Care Practitioner needs to be
aware of this effect and monitor the patient closely for any adverse side
effects.
2. Fill
medication nebulizer with Normal Saline along with prescribed dosage of
respiratory medication if indicated, for a total volume of 20 cc’s. (I.e. 0.5 cc Albuterol and 3.0 cc Mucomyst and 16.5
cc’s Normal Saline).
3. Place
ventilator Intrapulmonary Percussive Ventilator circuit (see drawing) inline.
A. A starting drive pressure of 30 psi is
recommended for ventilated patients.
B. A starting percussive rate of
approximately 200 bpm is recommended.
4. Remote
Switch knob must be activated manually. Percussion should be continued through
two complete ventilator cycles. Release Control knob and allow ventilator to
deliver severa1 machine breaths. Fine-tune and observe visible chest movement.
Monitor breath sounds and observe pulse oximeter for oxygen saturation
improvement.
5. After
several percussion intervals lower the endotracheal tube/tracheostomy tube cuff
pressure approximately 1-2 cmH2O to facilitate secretion
mobilization around the ET tube and to prevent possible ET tube obstruction by
secretion impaction.
6. If
chest percussion is inadequate, raise Drive Pressure (PSI gauge) and scan
Frequency Rate to mobilize secretions. Suction as required.
7. Therapy should continue
intermittently for approximately twenty minutes.
8. Patients
who are performing T-tube trials or CPAP sprinting may be taken off of the
ventilator for the Intrapulmonary Percussive Ventilation treatment utilizing an
elbow adapter and 6 inches of aerosol tubing.
Decreasing cuff pressure still applies to this patient population.
Note:
manually cancel occasional high-pressure alarms as they occur. The goal is to achieve approximately twice
the
9. At
Completion of Therapy
A.
Reset
cuff pressure.
B.
Remove
Intrapulmonary Percussive Ventilator circuit assembly and store appropriately.
C. Return
ventilator settings if changed prior to starting Intrapulmonary Percussive
Ventilation treatment.
INTRAPULMONARY
PERCUSSIVE VENTILATION (IPV)
SUGGESTED
VENTILATOR PROTOCOL
VENTILATOR
PATIENT PROCEDURE:
1. For
ALL BIRD ventilators, settings need to be set as follows: Treatment can be
done in either volume or pressure control
D.
In
the SIMV or CPAP Mode.
E.
Turn
off Pressure Support
F.
PEEP
levels may be maintained
1. Mean
Airway Pressures will increase slightly with the administration of
Intrapulmonary Percussive Ventilation inline with the ventilator. The Respiratory Care Practitioner needs to be
aware of this effect and monitor the patient closely for any adverse side
effects.
2. Fill
medication nebulizer with Normal Saline along with prescribed dosage of
respiratory medication if indicated, for a total volume of 20 cc’s. (I.e. 0.5 cc Albuterol and 3.0 cc Mucomyst and 16.5
cc’s Normal Saline).
3. Place
ventilator Intrapulmonary Percussive Ventilator circuit (see drawing) inline.
C. A starting drive pressure of 30 psi is
recommended for ventilated patients.
D. A starting percussive rate of
approximately 200 bpm is recommended.
4. Remote
Switch knob must be activated manually. Percussion should be continued through
two complete ventilator cycles. Release Control knob and allow ventilator to
deliver severa1 machine breaths. Fine-tune and observe visible chest movement.
Monitor breath sounds and observe pulse oximeter for oxygen saturation
improvement.
5. After
several percussion intervals lower the endotracheal tube/tracheostomy tube cuff
pressure approximately 1-2 cmH2O to facilitate secretion
mobilization around the ET tube and to prevent possible ET tube obstruction by
secretion impaction.
6. If
chest percussion is inadequate, raise Drive Pressure (PSI gauge) and scan
Frequency Rate to mobilize secretions. Suction as required.
7. Therapy should continue
intermittently for approximately twenty minutes.
8. Patients
who are performing T-tube trials or CPAP sprinting may be taken off of the
ventilator for the Intrapulmonary Percussive Ventilation treatment utilizing an
elbow adapter and 6 inches of aerosol tubing.
Decreasing cuff pressure still applies to this patient population.
Note:
manually cancel occasional high-pressure alarms as they occur. The goal is to achieve approximately twice
the
9. At
Completion of Therapy
C.
Reset
cuff pressure.
D.
Remove
Intrapulmonary Percussive Ventilator circuit assembly and store appropriately.
C. Return
ventilator settings if changed prior to starting Intrapulmonary Percussive
Ventilation treatment.
INTRAPULMONARY
PERCUSSIVE VENTILATION (IPV)
SUGGESTED
VENTILATOR PROTOCOL
VENTILATOR
PATIENT PROCEDURE:
1. For
Bear ventilators, settings need to be set as follows: Treatment can be done
in either volume or pressure control
G.
In
the SIMV or CPAP Mode.
H.
Turn
off Pressure Support
I.
PEEP
levels may be maintained
1. Mean
Airway Pressures will increase slightly with the administration of
Intrapulmonary Percussive Ventilation inline with the ventilator. The Respiratory Care Practitioner needs to be
aware of this effect and monitor the patient closely for any adverse side
effects.
2. Fill
medication nebulizer with Normal Saline along with prescribed dosage of
respiratory medication if indicated, for a total volume of 20 cc’s. (I.e. 0.5 cc Albuterol and 3.0 cc Mucomyst and 16.5
cc’s Normal Saline).
3. Place
ventilator Intrapulmonary Percussive Ventilator circuit (see drawing) inline.
E. A starting drive pressure of 30 psi is
recommended for ventilated patients.
F. A starting percussive rate of
approximately 200 bpm is recommended.
4. Remote
Switch knob must be activated manually. Percussion should be continued through
two complete ventilator cycles. Release Control knob and allow ventilator to
deliver severa1 machine breaths. Fine-tune and observe visible chest movement.
Monitor breath sounds and observe pulse oximeter for oxygen saturation
improvement.
5. After
several percussion intervals lower the endotracheal tube/tracheostomy tube cuff
pressure approximately 1-2 cmH2O to facilitate secretion
mobilization around the ET tube and to prevent possible ET tube obstruction by
secretion impaction.
6. If
chest percussion is inadequate, raise Drive Pressure (PSI gauge) and scan
Frequency Rate to mobilize secretions. Suction as required.
7. Therapy should continue
intermittently for approximately twenty minutes.
8. Patients
who are performing T-tube trials or CPAP sprinting may be taken off of the
ventilator for the Intrapulmonary Percussive Ventilation treatment utilizing an
elbow adapter and 6 inches of aerosol tubing.
Decreasing cuff pressure still applies to this patient population.
Note:
manually cancel occasional high-pressure alarms as they occur. The goal is to achieve approximately twice
the
9. At
Completion of Therapy
E.
Reset
cuff pressure.
F.
Remove
Intrapulmonary Percussive Ventilator circuit assembly and store appropriately.
C. Return
ventilator settings if changed prior to starting Intrapulmonary Percussive
Ventilation treatment.
INTRAPULMONARY
PERCUSSIVE VENTILATION (IPV)
SUGGESTED
VENTILATOR PROTOCOL
VENTILATOR
PATIENT PROCEDURE:
1. For
Servo 300A ventilators, settings need to be set as follows: Treatment can be
done in either volume or pressure control
J.
In
the SIMV or CPAP Mode.
K.
Turn
off Pressure Support
L.
PEEP
levels may be maintained
1.
Mean
Airway Pressures will increase slightly with the administration of
Intrapulmonary Percussive Ventilation inline with the ventilator. The Respiratory Care Practitioner needs to be
aware of this effect and monitor the patient closely for any adverse side
effects.
C. If patient has Automode on, it needs to
be turned off.
D. Treatment can be done in both PRVC and VS
modes of ventilation.
1.
In
PRVC and VS modes, when IPV is started and extra flow begins to be added to the
circuit, the 300A ventilator will make adjustment.
a.
Initially,
Vt will drop due to extra flow being added.
b.
The
300A will the increase pressure in order to compensate for the lost volume
Vt will be maintained during the rest of the treatment while
IPV is being delivered
c.
At
the end of the IPV treatment and it is turned off, the ventilator will deliver
1 – 4 breaths at higher pressures, which will result in a higher Vt being delivered to your patient.
d.
The
300A ventilator will, after approximately 4 breaths, return pressure to where
it was prior to IPV treatment
2. Fill
medication nebulizer with Normal Saline along with prescribed dosage of
respiratory medication if indicated, for a total volume of 20 cc’s. (i.e. 0.5 cc Albuterol and 3.0 cc Mucomyst and 16.5
cc’s Normal Saline).
3. Place
ventilator Intrapulmonary Percussive Ventilator circuit (see drawing) inline.
G. A starting drive pressure of 30 psi is
recommended for ventilated patients.
H. A starting percussive rate of
approximately 200 bpm is recommended.
4. Remote
Switch knob must be activated manually. Percussion should be continued through
two complete ventilator cycles. Release Control knob and allow ventilator to
deliver severa1 machine breaths. Fine-tune and observe visible chest movement.
Monitor breath sounds and observe pulse oximeter for oxygen saturation
improvement.
5. After
several percussion intervals lower the endotracheal tube/tracheostomy tube cuff
pressure approximately 1-2 cmH2O to facilitate secretion
mobilization around the ET tube and to prevent possible ET tube obstruction by
secretion impaction.
6. If
chest percussion is inadequate, raise Drive Pressure (PSI gauge) and scan
Frequency Rate to mobilize secretions. Suction as required.
7. Therapy should continue
intermittently for approximately twenty minutes.
8. Patients
who are performing T-tube trials or CPAP sprinting may be taken off of the
ventilator for the Intrapulmonary Percussive Ventilation treatment utilizing an
elbow adapter and 6 inches of aerosol tubing.
Decreasing cuff pressure still applies to this patient population.
Note:
manually cancel occasional high-pressure alarms as they occur. The goal is to achieve approximately twice
the
9. At
Completion of Therapy
G.
Reset
cuff pressure.
H.
Remove
Intrapulmonary Percussive Ventilator circuit assembly and store appropriately.
C. Return
ventilator settings if changed prior to starting Intrapulmonary Percussive
Ventilation treatment.
Considerations
For Delivering IPV In-Line In Any Pressure Mode Of
Ventilation
For IPV
treatments given in pressure control mode, when IPV is turned on, it adds flow
to the circuit. The ventilator senses
this extra flow and decreases the flow it has been delivering. This will result in a slight loss of
If the Vt being lost during the IPV treatment is consider significant,
you can overcome this problem by simply increasing your set PC limit to achieve
a desired
IPV Protocol Algorithm
